Innovative Methodology
School of Applied and Engineering Physics, Cornell University, Ithaca, New York 14853
Submitted 20 October 2003; accepted in final form 1 December 2003
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
Although routine use of multiphoton microscopy achieves imaging depths in brain of typically <500 µm, the ultimate imaging depth is dependent on many parameters, including the age of the animal and the optical geometry (Oheim et al. 2001
). Oheim et al. (2001
) have extended the reach of multiphoton microscopy by approximately 100 µm by appropriate use of a low-magnification, high numerical aperture objective. Theer et al. (2003
) recently used regenerative amplification of 200-kHz pulses to achieve high pulse peak powers while maintaining reasonable average powers, enabling imaging depths of close to 1 mm, although image quality suffered significantly at greater depths. Gradient index (GRIN) lenses offer the opportunity to reach even deeper lying structures.
GRIN lenses use a negative gradient in the refractive index of glass from the center of the lens to the outside edge to bend and focus light. GRIN lenses are characterized by a length, or pitch, and a numerical aperture (NA). The pitch of a GRIN lens determines how many internal images are formed within the lens. A 0.25-pitch lens focuses a parallel beam incident on the front surface of the lens to a point on the back surface. A lens of pitch 1 forms an upright image on the back surface, with an internal, inverted image plane located at one-half the length of the lens. GRIN lenses are commercially available in lengths of up to several tens of centimeters, long enough, in principle, to access deep brain structures in large animals and humans.
GRIN lenses have been used for fiber bundlecoupled confocal microscopy (Knittel et al. 2001
) and in vivo epi-fluorescence microendoscopy (Jung and Schnitzer 2002
). The use of large (>1 mm diam) GRIN lenses for in vivo multiphoton microscopy (Levene et al. 2002
) and the characterization of thin composite GRIN lenses for multiphoton microscopy (Jung and Schnitzer 2003
) have recently been reported. Multiphoton microscopy (Denk et al. 1990
) has several important advantages for in vivo imaging applications with GRIN lenses. It is capable of resolving optical sections relatively far from the surface of the lens, while tissue in close proximity to the lens may suffer from mechanical damage or immune response. The superior sectioning ability of multiphoton excitation is critical for observing detailed structure in tissues in which fluorescence sources are distributed throughout the region of interest. Multiphoton microscopy is able to excite intrinsic tissue fluorescence and UV absorbing dyes without the use of UV excitation, which has poor tissue penetration and produces a strong fluorescent background both from tissue regions outside the focal plane and from the high numerical aperture GRIN lenses themselves.
| METHODS |
|---|
|
|
|---|
The multiphoton microscope used here is the same as described elsewhere (Kloppenburg et al. 2000
) except that, for yellow fluorescent protein (YFP) imaging, the bialkali photomultiplier tube (PMT) is replaced by a GaAsP PMT (H7422P, Hamamatsu, Bridgewater, NJ), which exhibits a fivefold greater sensitivity to yellow wavelengths. The Ti:Sapphire laser source produced
100-fs pulses at 80 MHz. The position of the GRIN lens relative to the objective lens is controlled via hydraulic micropositioners (Narishige Scientific Instruments Lab, Tokyo, Japan) fixed to the nosepiece of the microscope. The focus knob of the microscope moves the objective and GRIN lens together as a fixed unit to the appropriate depth within the specimen. A piezoelectric focus controller (Physik Instrumente, Waldbronn, Germany) adjusts the focus of the objective relative to the GRIN lens, allowing z-series of the sample while holding the position of the GRIN lens fixed.
Animal surgeries
Anesthesia and surgeries were performed in accordance with Cornell Universityapproved animal use protocols. The mice were maintained under ketamine (76 mg/kg)/xylazine (5 mg/kg) anesthesia during the surgery and the following imaging session. To reveal the cortex for GRIN rod penetration, we used a dental drill to create a circular craniotomy (5 mm diam) centered above the dorsal parietal cortex. The dura visible beneath was cut at the edge of the field and carefully removed. For analgesia at the site of the craniotomy, lidocaine (2%) was applied topically. The body temperature was maintained at
36° celsius using a feedback-controlled heating pad during both surgery and imaging. Prior to the microangiography, the blood serum was labeled by a tail-vein injection of a 100-µl bolus of either fluorescein-labeled dextran (20 mg/ml, 500 kDa, Molecular Probes) or quantum dots (35 µM, 608 nm emission, Quantum Dot, Hayward, CA) in physiological saline.
GRIN lens and imaging parameters
The use of GRIN lenses for in vivo multiphoton microscopy is shown in Fig. 1. One end of the GRIN lens is positioned close to the focal plane of the objective lens of a standard multiphoton laser scanning microscope, with the opposite end inserted inside the specimen under study. The GRIN lens refocuses the laser light tens of micrometers from the opposite end of the lens, inside the specimen, where the scanned beam excites fluorescence that is collected back through the GRIN lens and the objective. Dispersion compensation is not required because pulse dispersions in the GRIN lenses used here are
2,700 fs2, similar to typical high-NA objective lenses.
|
16 mm. All but the last 900 µm at either end of the composite GRIN lens was protected from mechanical damage by a metal sheath with an outside diameter of 600 µm. While high NA end pieces determine the resolution of the composite lens, it was necessary to use lower NA material for the central length of lens to avoid focusing the excitation laser pulse inside the high NA material, which can produce a fluorescent background and may result in nonlinear optical effects such as self-phase modulation. This choice, however, reduces the total field of view to a 58-µm-diam circle. Other configurations may be used when larger fields of view are required. The pitch of the 0.6-NA material was chosen to give a focal distance of 35 µm in air and 47 µm in water. Changing the position of the focal plane of the objective results in a change in the working distance (WD) from the end of the GRIN lens in the sample. This allows z-scanning in the sample, from 0 to 95 µm from the surface of the lens, without moving the GRIN lens.
|
66%, resulting in an overall excitation efficiency of
40%. The full width half-maximum (FWHM) of the lateral point spread function (PSF), measured using sub-resolution fluorescent particles, was 825 nm, less than twice the diffraction limited value of 481 nm and similar to previous reports (Jung and Schnitzer 2003
In all cases, the GRIN lens was inserted approximately perpendicularly to the cortical surface through the cranial opening in steps of a few tens of micrometers. Pausing between steps allowed tissue to physically readjust to the presence of the lens. It has been suggested that rapid penetration by blunt tips in neural tissue minimizes tissue damage, while slow penetration minimizes tissue compression (Edell et al. 1992
). Pausing between rapid steps should therefore minimize both tissue damage and compression. Care was taken to avoid inserting the lens at locations with considerable blood from surgery on the cortical surface. Once inserted, most trials produced useful images, with little, if any, apparent accumulation of blood on the lens surface during or after penetration.
For the images of blood vessels, the excitation wavelength was 830 nm in Fig. 3, AC, and 780 nm in Fig. 3D, with
50 mW of power at the sample. Figure 3, AC, shows 1-s scans taken at a WD of 50 µm. Figure 3D is a projection of five images taken with WDs of 1340 µm from the end of the GRIN lens. Each image in the stack is the average of two scans lasting 3 s each.
|
|
| RESULTS AND DISCUSSION |
|---|
|
|
|---|
800 µm below the surface of wild-type mouse cortex after administration of fluorescent quantum dots by tail vein injection. Fluorescent quantum dots have high two-photon cross-sections and are excellent indicators for in vivo angiography (Larson et al. 2003
0.6 mm/s, similar to the velocity measured in rat cortex (Chaigneau et al. 2003
5-µm-diam capillary appears on the left, while a larger,
15-µm-diam blood vessel is on the right. Although blood flow was observed, the orientation of the capillary was not conducive to velocity estimates using the line scan technique. Previous attempts at fluorescence microangiography in brain were limited to vessels <600 µm below the surface of cortex (Kleinfeld et al. 1998
THY1-YFP line H mice (Jackson Laboratory, Bar Harbor, ME) express YFP in cortical layer V neurons and in the hippocampus (Feng et al. 2000
), making them ideal candidates for deep in vivo imaging. Figure 4, AC, shows layer V neuron cell bodies 700800 µm below the surface of the cortex and represents the typical range of image quality in cortex for these lenses. Axonal bundles in the subcortical white matter
1 mm below the surface were also visible (Fig. 4D). Figure 4E is an image of neuropil from the CA1 region of the hippocampus
1.5 mm below the surface of cortex. Figure 4, AD, demonstrates subcellular resolution, with clearly visible individual neuronal processes.
There are three primary aspects of possible tissue trauma due to GRIN lens penetration in solid organs such as the brain. First, the inevitable destruction of tissue along the penetration path with possible bleeding, tissue edema, and elevated intracranial pressure can be potential complications. Second, the compression of tissue of interest at the front end of the lens, and third, the subsequent formation of scar tissue and inflammation must also be considered.
The maximal thickness of the penetration path of 600 µm is well within the range of established acute and chronic brain implants such as push-pull cannulae (Philippu 1984
). By avoiding major veins along the perforation path, it was possible to achieve penetration such that bleeding was not detected at the front surface of the lens. However, cleavage of blood vessels along the path of penetration is inevitable and must be taken into account when interpreting the results of angiography.
Parietal cortex tolerated the implantation of the GRIN lens without clinical signs of elevated intracranial pressure. However, regions with closer proximity to the brain stem (occipital cortex, cerebellum) may require caution to avoid transtentorial herniation. The penetration depth of multiphoton microscopy allows imaging
60 µm from the end of the GRIN lens, where the inevitable tissue compression is substantially alleviated. Accordingly, we did not observe gross morphological changes indicating compression in our images. Although blood flow velocity can vary greatly even within the same region of cortex (Chaigneau et al. 2003
), our measured velocity of 0.6 mm/s was within the expected range of values and is therefore consistent with a lack of tissue compression. The accumulation of compressed tissue at the front of the GRIN lens during insertion was observed to be limited to
10 µm and did not interfere significantly with image acquisition.
It is likely that chronic implantation of GRIN lenses may induce gliosis and inflammation; however, studies of gliosis in response to electrode penetrations found responses limited to within the WD of the GRIN lenses (Edell et al. 1992
; Turner et al. 1999
) and independent of probe dimensions (Szarowski et al. 2003
). The feasibility of multiphoton microscopy in conjunction with cranial windows has been consistently demonstrated in chronic imaging studies (Bacskai et al. 2001
; Christie et al. 2001
).
In addition to choosing lenses of the smallest practicable diameter, the impact of tissue damage in the brain may be minimized by working in larger animals and by using the lenses to image in previously inaccessible sulci without actually penetrating brain tissue. Use of intrinsic signals (Dombeck et al. 2003
; Zipfel et al. 2003
) or alternative labeling methods, such as calcium indicators (Svoboda et al. 1997
) or dyes that stain
-amyloid plaques (Christie et al. 2001
), with this deep imaging technique could enable many new physiological experiments both in basic research and in animal models of neurodegenerative diseases. This imaging modality may also be easily combined with conventional in vivo electrophysiological techniques.
GRIN lenses used in conjunction with multiphoton microscopy open up the possibility of performing deep in vivo fluorescence imaging in brain. We have demonstrated the compatibility of GRIN lenses with in vivo imaging. Potential complications such as motion artifact and the accumulation of damaged tissue at the front of the lens during penetration have been shown to be negligible. Scanned fiber optic coupling to long lenses may eventually enable work in large animals and in awake behaving animals. Considerable progress in fiber optic scanning for in vivo imaging of awake behaving animals has already been made by Helmchen and co-workers (Helmchen et al. 2001
; Ouzounov et al. 2002
). We anticipate that in vivo multiphoton imaging with GRIN lenses will prove to be a valuable tool for both biological research and clinical applications.
| ACKNOWLEDGMENTS |
|---|
|
|
|---|
GRANTS
This work was supported by the facilities of the National Institutes of Health (NIH) Biomedical Resource funded by the National Center for Research Resources-NIH/National Institute of Biomedical Imaging and Bioengineering. Additional support was received from the U.S. Department of Energy, the National Science Foundation, and from an NIH Training Grant in Molecular Biophysics.
| FOOTNOTES |
|---|
Address for reprint requests and other correspondence: W. W. Webb, Clark Hall, Rm. 223, Cornell Univ., Ithaca, NY 14853 (E-mail: www2{at}cornell.edu).
| REFERENCES |
|---|
|
|
|---|
Chaigneau E, Oheim M, Audinat E, and Charpak S. Two-photon imaging of capillary blood flow in olfactory bulb glomeruli. Proc Natl Acad Sci USA 100: 1308113086, 2003.
Christie RH, Bacskai BJ, Zipfel WR, Williams RM, Kajdasz ST, Webb WW, and Hyman BT. Growth arrest of individual senile plaques in a model of Alzheimer's disease observed by in vivo multiphoton microscopy. J Neurosci 21: 858864, 2001.
Denk W, Strickler JH, and Webb WW. Two-photon laser scanning fluorescence microscopy. Science 248: 7376, 1990.
Dombeck DA, Kasischke KA, Vishwasrao HD, Ingelsson M, Hyman BT, and Webb WW. Uniform polarity microtubule assemblies imaged in native brain tissue by second-harmonic generation microscopy. Proc Natl Acad Sci USA 100: 70817086, 2003.
Edell DJ, Toi VV, McNeil VM, and Clark LD. Factors influencing the biocompatibility of insertable silicon microshafts in cerebral cortex. IEEE Trans. Biomed. Eng. 39: 635643, 1992.[CrossRef][ISI][Medline]
Feng G, Mellor RH, Bernstein M, Kellr-Peck C, Nguyen QT, Wallace M, Nerbonne JM, Lichtman JW, and Sanes JR. Imaging neuronal subsets in transgenic mice expressing multiple spectral variants of GFP. Neuron 28: 4151, 2000.[CrossRef][ISI][Medline]
Helmchen F, Fee MS, Tank DW, and Denk W. A miniature head-mounted two-photon microscope: high-resolution brain imaging in freely moving animals. Neuron 31: 903912, 2001.[CrossRef][ISI][Medline]
Jacobs RE and Cherry SR. Complementary emerging techniques: high resolution PET and MRI. Curr Opin Neurobiol 11: 621629, 2001.[CrossRef][ISI][Medline]
Jung JC and Schnitzer MJ. In vivo epi-fluorescence micro-edoscopy. Biophys J 82: 496a, 2002.
Jung JC and Schnitzer MJ. Multiphoton endoscopy. Optics Lett 28: 902904, 2003.[ISI][Medline]
Kleinfeld D, Mitra PP, Helmchen F, and Denk W. Fluctuations and stimilus-induced changes in blood flow observed in individual capillaries in layers 2 through 4 of rat neocortex. Proc Natl Acad Sci USA 95: 1574115746, 1998.
Kloppenburg P, Zipfel WR, Webb WW, and Harris-Warrick RM. Highly localized Ca2+ accumulation revealed by multiphoton microscopy in an identified motoneuron and its modulation by dopamine. J Neurosci 20: 25232533, 2000.
Knittel J, Schnieder L, Buess G, Messerschmidt B, and Possner T. Endoscope-compatible confocal microscope using a gradient index-lens system. Optic Comm. 188: 267273, 2001.[CrossRef]
Larson DR, Zipfel WR, Williams RM, Clark SW, Bruchez MP, Wise FW, and Webb WW. Water-soluble quantum dots for multiphoton fluorescence imaging in vivo. Science 300: 14341436, 2003.
Levene MJ, Dombeck DA, Kasischke KA, Molloy RP, and Webb WW. Multi-photon imaging of deep tissue using gradient index lenses. Society for Neuroscience Annual Meeting, Orlando, FL, Nov. 27, 2002.
Oheim M, Beaurepaire E, Chaigneau E, Mertz J, and Charpak S. Two-photon microscopy in brain tissue: parameters influencing the imaging depth. J Neurosci Methods 111: 2937, 2001.[CrossRef][ISI][Medline]
Ouzounov DG, Moll KD, Foster MA, Zipfel WR, Webb WW, and Gaeta AL. Delivery of nanojoule femtosecond pulses through large-core microstructured fibers. Optics Lett 27: 15131515, 2002.[Medline]
Philippu A. Use of push-pull cannulae to determine the release of endogenous neurotransmitters in distinct brain areas of anaesthetized and freely moving animals. In: Measurement of neurotransmitter release in vivo, edited by Marsden CA. New York: John Wiley, 1984, p. 337.
Svoboda K, Denk W, Kleinfeld D, and Tank DW. In vivo dendritic calcium dynamics in neocortical pyramidal neurons. Nature 385: 161165, 1997.[CrossRef][Medline]
Szarowski DH, Andersen MD, Retterer S, Spence AJ, Isaacson M, Craighead H, Turner JN, and Shain W. Brain response to micro-machined silicon devices. Brain Res 983: 2335, 2003.[CrossRef][ISI][Medline]
Theer P, Hasan MT, and Denk W. Two-photon imaging to a depth of 1000 µm in living brains by use of a Ti:Al2O3 regenerative amplifier. Optics Lett 28: 10221024, 2003.[ISI][Medline]
Turner JN, Shain W, Szarowski DH, Andersen M, Martins S, Isaacson M, and Craighead H. Cerebral astrocyte response to micromachined silicon implants. Exp Neurol 156: 3349, 1999.[CrossRef][ISI][Medline]
Zipfel WR, Williams RW, Christie R, Nikitin AY, Hyman BT, and Webb WW. Live tissue intrinsic emission microscopy using multiphoton-excited native fluorescence and second harmonic generation. Proc Natl Acad Sci USA 100: 70757080, 2003.
This article has been cited by other articles:
![]() |
F. Legare, C. Pfeffer, and B. R. Olsen The Role of Backscattering in SHG Tissue Imaging Biophys. J., August 15, 2007; 93(4): 1312 - 1320. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Fu, H. Wang, R. Shi, and J.-X. Cheng Second Harmonic and Sum Frequency Generation Imaging of Fibrous Astroglial Filaments in Ex Vivo Spinal Tissues Biophys. J., May 1, 2007; 92(9): 3251 - 3259. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Deisseroth, G. Feng, A. K. Majewska, G. Miesenbock, A. Ting, and M. J. Schnitzer Next-Generation Optical Technologies for Illuminating Genetically Targeted Brain Circuits J. Neurosci., October 11, 2006; 26(41): 10380 - 10386. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Yao, D. R. Larson, H. D. Vishwasrao, W. R. Zipfel, and W. W. Webb Blinking and nonradiant dark fraction of water-soluble quantum dots in aqueous solution PNAS, October 4, 2005; 102(40): 14284 - 14289. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Wang, Y. Fu, P. Zickmund, R. Shi, and J.-X. Cheng Coherent Anti-Stokes Raman Scattering Imaging of Axonal Myelin in Live Spinal Tissues Biophys. J., July 1, 2005; 89(1): 581 - 591. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Zenisek Vesicle reuse revisited PNAS, May 24, 2005; 102(21): 7407 - 7408. [Full Text] [PDF] |
||||
![]() |
E. C. Rothstein, S. Carroll, C. A. Combs, P. D. Jobsis, and R. S. Balaban Skeletal Muscle NAD(P)H Two-Photon Fluorescence Microscopy In Vivo: Topology and Optical Inner Filters Biophys. J., March 1, 2005; 88(3): 2165 - 2176. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. E. Stutzmann and I. Parker Dynamic Multiphoton Imaging: A Live View from Cells to Systems Physiology, February 1, 2005; 20(1): 15 - 21. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Rubart Two-Photon Microscopy of Cells and Tissue Circ. Res., December 10, 2004; 95(12): 1154 - 1166. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Jung, A. D. Mehta, E. Aksay, R. Stepnoski, and M. J. Schnitzer In Vivo Mammalian Brain Imaging Using One- and Two-Photon Fluorescence Microendoscopy J Neurophysiol, November 1, 2004; 92(5): 3121 - 3133. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Brecht, M. S. Fee, O. Garaschuk, F. Helmchen, T. W. Margrie, K. Svoboda, and P. Osten Novel Approaches to Monitor and Manipulate Single Neurons In Vivo J. Neurosci., October 20, 2004; 24(42): 9223 - 9227. [Full Text] [PDF] |
||||
![]() |
S. X. Lin and F. R. Maxfield Fluorescence imaging in living animals. Focus on "Uptake and trafficking of fluorescent conjugates of folic acid in intact kidney determined using intravital two-photon microscopy" Am J Physiol Cell Physiol, August 1, 2004; 287(2): C257 - C259. [Full Text] [PDF] |
||||
![]() |
A. Mizrahi, J. C. Crowley, E. Shtoyerman, and L. C. Katz High-Resolution In Vivo Imaging of Hippocampal Dendrites and Spines J. Neurosci., March 31, 2004; 24(13): 3147 - 3151. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |